16.3.6
"Maintenance" events related to calibration (Orange LED B
(right) and icons
When a "maintenance" event related to calibration is generated:
• The 4...20 mA output(s) operate normally.
• The transistor outputs configured in "warning" mode switch.
Message dis-
played in the
"Information"
menu
"MxM:Time to
cal."
"W:AI1 cal."
"W:AI2 cal."
"W:AI1 maint."
"W:AI2 maint."
→ To get information on a message which is not listed, contact Bürkert.
208
,
Meaning
A calibration is due on the sensor in the "Mx"
module.
The periodicity of calibrations is set in the
"INTERVAL" function in the "CALIBRATION
INTERVAL" menu (see section 10, page 148).
A calibration is due for AI1.
The periodicity of the calibrations is set in
chapter 10.13.
A calibration is due for AI2.
The periodicity of the calibrations is set in
chapter 10.13.
A maintenance operation is due for AI1.
The periodicity of the maintenance operations is
set in chapter 10.15.
A maintenance operation is due for AI2.
The periodicity of the maintenance operations is
set in chapter 10.15.
English
and
displayed)
Recommended action
→ Calibrate the sensor (section 10,
→ Do a calibration of AI1. Refer to
→ Reset the delay until the next cali-
→ Do a calibration of AI2. Refer to
→ Reset the delay until the next cali-
→ Do the necessary maintenance
→ Reset the delay until the next
→ Do the necessary maintenance
→ Reset the delay until the next
Type 8619
Repair and maintenance
page 148).
chapters 10.5 to 10.10.
bration. Refer to chapter 10.13.
chapters 10.5 to 10.10.
bration. Refer to chapter 10.13.
operation on AI1.
maintenance operation. Refer to
chapter 10.15.
operation on AI2.
maintenance operation. Refer to
chapter 10.15.