Setting The Parameters Of A Ph/Redox Module - Burkert 8619 Operating Instructions Manual

Modular transmitter/controller
Hide thumbs Also See for 8619:
Table of Contents

Advertisement

9.28

Setting the parameters of a pH/redox module

Refer to section 8.10 to access the "Parameters" menu.
Parameters
Mx:pH/ORP
This is
when the
device is be-
ing parame-
tered............
....................
RTD: Choose the type of temperature sensor connected to the module.
TEMPERATURE: Choose the value of the temperature used in the process (particularly to compensate for
the pH measurement):
• choose "Auto": the temperature of the fluid is measured by the sensor.
• choose "Manual": enter the value of the process temperature (in °C) in the next field, e.g. when no tem-
perature sensor is connected to the module.
ADJUST TEMP.: The measured temperature can be corrected by an offset value. Enter the offset value in °C.
TEMP. CALIBRATION: Choose the value of the temperature used when calibrating the probe/sensor:
• choose "Auto": the temperature of the fluid is measured by the sensor.
• choose "Manual": enter the value of the calibration temperature (in °C) in the next field, e.g. when no tem-
perature sensor is connected to the module.
142
RTD
This is
when the
device is be-
ing parame-
tered............
....................
Temperature
Temp. adjust
°C
Temp. calib
°C
Buffer
Calibration limits
English
None
PT100
PT1000
Auto
Manual
if "Temperature" = "Auto"
ENTERING
if "Temperature" = "Manual"
ENTERING
Auto
Manual
if "Temp. calib" = "Manual"
ENTERING
Hamilton
DIN19267
pH zero
Warning high:
Warning low:
Error high:
Error low:
pH slope
Warning high:
Warning low:
Error high:
Error low:
ORP Offset
Warning high:
Warning low:
Error high:
Error low:
Type 8619
Parameters menu
ENTERING
ENTERING
ENTERING
ENTERING
ENTERING
ENTERING
ENTERING
ENTERING
ENTERING
ENTERING
ENTERING
ENTERING