Monitoring The Temperature Of The Fluid - Burkert 8619 Operating Instructions Manual

Modular transmitter/controller
Hide thumbs Also See for 8619:
Table of Contents

Advertisement

• The "warning" event may also be associated with one and/or other digital outputs. See section 9.26.
• A current of 22 mA may be emitted on one and/or other of the current outputs when an "error" event
related to either the monitoring of the fluid pH, redox, conductivity or temperature values or the moni-
toring of an analogue input is generated. See section 9.25.
• See also section 16.3 If you encounter problems.
STATE: Choose whether or not to activate monitoring of fluid conductivity.
This monitoring is done by the generation of a "warning" event if the fluid conductivity range defined in the
"Warn Hi/Lo" functions below is exceeded and an "error" event if the fluid conductivity range defined in the
"Err Hi/Lo" functions below is exceeded.
CONDUCTIVITY: Read the fluid conductivity measured in real time by the sensor.
WARN. HIGH: Enter the fluid conductivity value above which a "warning" event is generated.
WARN. LOW: Enter the fluid conductivity value below which a "warning" event is generated.
ERR. HIGH: Enter the fluid conductivity value above which an "error" event is generated.
ERR. LOW: Enter the fluid conductivity value below which an "error" event is generated.
11.7

Monitoring the temperature of the fluid

This function is used to monitor the fluid conductivity and define the device's behaviour if the ranges
defined are exceeded.
Refer to section 8.10 to access the "Diagnostics" menu.
Diagnostics
Mx:pH/ORP
Mx:Conductivity
A malfunction in your process or on the temperature sensor may be revealed by too low or too high a fluid
temperature or by an incorrect temperature measurement.
To be warned when the temperature measurement is out of range:
→ activate monitoring on the fluid temperature in the "Status" function, and then
→ set a temperature range (in °C) outside of which the 8619 generates a "warning" event and displays the
icons
and
→ set a temperature range (in °C) outside of which the 8619 generates an "error" event and displays the
icons
and
When a "warning" or "error" event is generated by the 8619:
→ go into the "Information" menu to read the cause of the event generation
→ and/or read the measured temperature value.
→ check whether the temperature sensor is working correctly by measuring a fluid with a known temper-
ature. If the temperature sensor is faulty, return the device to Bürkert;
→ if the temperature sensor is not the cause of the problem, check the process.
172
Temperature
;
X
.
English
State:
ON/OFF
Temperature
READING
Warning high:
ENTERING
Warning low:
ENTERING
Error high:
ENTERING
Error low:
ENTERING
Type 8619
Diagnostics Menu