Burkert 8619 Operating Instructions Manual page 146

Modular transmitter/controller
Hide thumbs Also See for 8619:
Table of Contents

Advertisement

ALARM USP: Enter a conductivity value percentage from the table "USP <645>".
χ
Table 10: Behaviour associated to the table USP<645>
Graph
Description
zone
C
The conductivity of the fluid has
exceeded the value in the table
USP<645> at the corresponding
temperature.
B
The conductivity of the fluid is
between the percentage set in
the function "USP Alarm" and the
value in the table USP<645> at
the corresponding temperature.
A
The conductivity of the fluid is
below the percentage set in the
function "USP Alarm" at the
corresponding temperature.
CALIB. TEMP. : Choose the value of the temperature used when calibrating the probe/sensor:
• choose "Auto": the temperature of the fluid is measured by the sensor
• choose "Manual": enter the value of the temperature (in °C) in the next field, e.g. when no temperature
sensor is connected to the module.
CALIBRATION SOLUTION : Choose the calibration solution used for automatic calibration of the conduc-
tivity sensor.
146
C
B
A
English
USP<645> graph
Percentage of the USP<645> graph, set in
the "USP Alarm" field
T [°C]
Name displayed in
Associated code
the user defined
in the datalogger
"Ux" view (see
(see section 9.22)
section 9.9)
"> Max."
1
"USP Alarm"
2
"OK"
0
Type 8619
Parameters menu
Status of the "ON/
OFF" output (see
section 9.26)
ON (output not
inverted)
ON (output not
inverted)
OFF (output not
inverted)