Specifications Of The "Ph/Orp" Module - Burkert 8619 Operating Instructions Manual

Modular transmitter/controller
Hide thumbs Also See for 8619:
Table of Contents

Advertisement

6.10.5

Specifications of the "pH/ORP" module

pH measurement
• pH measurement range
• Resolution of pH measurement
• Systematic variation in the pH
measurement
• Potential difference measurement range
• Resolution of the potential difference
measurement
• Systematic variation in the potential dif-
ference measurement
• pH probe type
Power consumption
Measurement of the oxidation reduction
potential
• Oxidation reduction potential meas-
urement range
• Resolution of the potential difference
measurement
• Systematic variation in the potential dif-
ference measurement
• Oxidation reduction potential probe type
Temperature measurement
• Measurement range
• Measurement resolution
• Systematic variation in the measurement
• Temperature sensor type
34
English
• –2.00...+16.00 pH
• 0.01 pH
• ±0.02 pH + pH probe error
• –600...+600 mV
• 0.1 mV
• ±1 mV + pH probe error
• Electrochemical
0.1 VA
• –2000...+2000 mV
• 0.1 mV
• ±1 mV + ORP probe error
• Electrochemical
• –25...+130 °C
• 0.1 °C
• ±1 °C + temperature probe error
• Pt100 or Pt1000, with 2 or 3 wires
Type 8619
Technical data

Advertisement

Table of Contents
loading

This manual is also suitable for:

8619 multicell wm ac8619 multicell wm dc8619 multicell

Table of Contents