10
CALIBRATION MENU
10.1
Safety instructions
WARNING
Risk of injury due to non-conforming adjustment.
Non conforming adjustment could lead to injuries and damage the device and its environment.
▶ The operators in charge of adjustment must have read and understood the contents of this operating
instructions.
▶ In particular, observe the safety recommendations and intended use.
▶ The device/installation must only be adjusted by suitably trained staff.
10.2
Enabling/disabling the Hold function
Refer to section 8.10 to access the "Calibration" menu.
Calibration
System
The Hold mode is automatically deactivated when the device restarts after a power interruption, if
the Hold mode was activated at the moment of the power cut-off.
The Hold mode is used to carry out maintenance work without interrupting the process.
To activate the HOLD mode:
→ access the "HOLD" function;
→ choose "Enable";
→ confirm by "OK".
When the device is in Hold mode:
• the display shows the icon
• the current emitted on each 4...20 mA output is fixed at the last value of the process input associated with
each output;
• each digital output is fixed at the state acquired at the moment the Hold function is activated;
• the device remains in Hold mode until the Hold function is disabled.
To disable the HOLD mode:
→ access the "HOLD" function;
→ choose "Disable"
→ confirm by "OK".
148
Hold
H
instead of the icon
English
Disable
Enable
;
Type 8619
Calibration menu