Calibration menu
10.3

Modifying the Calibration menu access code

Refer to section 8.10 to access the "Calibration" menu. If the default access code "0000" is kept, the device
does not request it to access the Calibration menu.
Calibration
System
10.4

Adjusting the current outputs

Ensure that Hold mode is disabled before adjusting the current outputs: icon
display.
Refer to section 8.10 to access the "Calibration" menu.
Calibration
M0:Outputs
Mx:Outputs
4 MA: Adjust the offset of the current output.
When the "4mA" function is selected, the device generates a current of 4 mA:
→ measure the current emitted by the 4...20 mA output using a multimeter;
→ enter the value given by the multimeter.
20 MA: Adjust the span of the current output.
When the "20mA" function is selected, the device generates a current of 20 mA:
→ measure the current emitted by the 4...20 mA output using a multimeter;
→ enter the value given by the multimeter.
Code
0***
Enter the new
calibration menu
access code
AO1/AO2
4mA
20mA
Confirm code
0***
Confirm the new
code