Chapter 12: Troubleshooting the Emulation Probe

Problems with the Emulation Module

Problems with the Emulation Module
Occasionally you may suspect a hardware problem with the emulation module.
The procedure in this section describes how to test the emulation module
hardware, and if a problem is found, how to repair or replace the broken
component.
This procedure tests the hardware within the logic analysis system—the
emulation module and its connection to the logic analysis system. To test the
emulation probe, see page 177.

To test the emulation module

1 End any Emulation Control Interface or debugger sessions.
2 Disconnect the emulation probe from the target system.
3 In the system window, select the System Admin icon.
4 Select the Admin tab.
5 Select Self-Test....
6 Read the Question dialog and select Yes if you wish to run the self tests.
7 Select the Master Frame tab.
8 Select the E5901B Emulation Module.
9 Select Test All.
10 When you are finished running self tests, select Quit. Then restart your
session from the Session Manager window.