Installation Note
Extended Frequency Range Upgrade Kit (50 GHz to 67 GHz)
Upgrade Kit Number: E8364-60105
For E8364B/C Microwave Network Analyzers WITHOUT the
Configurable Test Set Option (Option 014)
Agilent Part Number: E8364-90022
Printed in USA December 2007
Supersedes Print Date: April 2007
© Agilent Technologies, Inc. 2003, 2004, 2007

Advertisement

loading

Summary of Contents for Agilent Technologies E8364-60105

  • Page 1 Extended Frequency Range Upgrade Kit (50 GHz to 67 GHz) Upgrade Kit Number: E8364-60105 For E8364B/C Microwave Network Analyzers WITHOUT the Configurable Test Set Option (Option 014) Agilent Part Number: E8364-90022 Printed in USA December 2007 Supersedes Print Date: April 2007 © Agilent Technologies, Inc. 2003, 2004, 2007...
  • Page 2: Warranty Statement

    “Restricted computer software” as defined in FAR 52.227-19 (June 1987) or any equivalent agency regulation or contract clause. Use, duplication or disclosure of Software is subject to Agilent Technologies’ standard commercial license terms, and non-DOD Departments and Agencies of the U.S. Government will receive no greater than Restricted Rights as defined in FAR 52.227-19(c)(1-2) (June 1987).
  • Page 3: Contacting Agilent

    Contacting Agilent Assistance with test and measurements needs and information on finding a local Agilent office are available on the Web at: www.agilent.com/find/assist In any correspondence or telephone conversation, refer to the Agilent NOTE product by its model number and full serial number. With this information, the Agilent representative can determine whether your product is still within its warranty period.
  • Page 4 Check the contents of your kit against the following list. If any item is missing or damaged, contact Agilent Technologies. Refer to “Contacting Agilent” on page Table 1 Option 067 Upgrade Kit (E8364-60105) to Extend the E8364B/C to 67 GHz Ref. Description Part Number Desig.
  • Page 5 Table 1 Option 067 Upgrade Kit (E8364-60105) to Extend the E8364B/C to 67 GHz Ref. Description Part Number Desig. 0515-1862 Machine screw M3.0 x 27.5 (for attaching E8361-20063 lid clamps to SOMA 70s) Lid clamp - SOMA 70s E8361-20063 Ribbon cable, A23/A24 SOMA 70s to A16 test set motherboard...
  • Page 6 Installation Procedure for the Upgrade Kit The network analyzer must be in proper working condition prior to installing this option. Any necessary repairs must be made before proceeding with this installation. This installation requires the removal of the analyzer’s protective outer WARNING covers.
  • Page 7 Overview of the Installation Procedure Step 1. Verify the Model, Serial, and Option Numbers of the Analyzer Step 2. Remove the Outer and Inner Covers Step 3. Replace the A12 Source 20 Step 4. Replace the Top Side to Bottom Side Cables Step 5.
  • Page 8 Test Equipment Required for Post-Upgrade Adjustments Description Agilent Part Number Power meter E4418B/E4419B Power sensor, Type-N (100 kHz–4.2 GHz) 8482A Power sensor, 2.4 mm (4.2 GHz–50 GHz) 8487A V8486A Power sensor, V-Band (50 GHz–67 GHz) V281B Adapter, WR-15 to 1.85 mm (m) 85058-60114 Adapter, 1.85 mm (f) to 1.85 mm (f) Adapter, 3.5 mm (m) to 2.4 mm (f)
  • Page 9 Step 2. Remove the Outer and Inner Covers Remove the Outer Cover This procedure is best performed with the analyzer resting on its front CAUTION handles in the vertical position. Do not place the analyzer on its front panel without the handles. This will damage the front panel assemblies. 1.
  • Page 10 Remove the Inner Cover 1. Place the analyzer top-side up on a flat surface. ① 2. Using a T-10 TORX driver, remove the eleven screws (item 3. Lift off the cover. Figure 3 Inner Cover Removal The figure above shows the E8364B front panel and floppy disk drive. The NOTE E8364C front panel has a slightly different appearance and does not include a floppy disk drive.
  • Page 11 Step 3. Replace the A12 Source 20 ① 1. With a 5/16-inch wrench, disconnect semirigid cable (item 2. Disengage the A12 source 20 from the A11 connector and slide it out of the analyzer while observing the below. CAUTION ① To avoid damage to the semirigid cable (item ), pull down on the cable from the CAUTION...
  • Page 12 3. Slide the new A12 source 20 into the analyzer while observing the above. Be sure CAUTION that the assembly is seated completely into the A11 connector. 4. Reconnect the semirigid cable and tighten with a 5/16-inch torque wrench. Installation Note E8364-90022...
  • Page 13 Step 4. Replace the Top Side to Bottom Side Cables 1. Place the analyzer on its side as shown. ① 2. Remove the cable hold-down wire (item 3. Disconnect cables W19 and W20 at both ends; two top side connections and two bottom side connections.
  • Page 14 Step 5. Raise the Receiver Deck 1. Place the analyzer bottom-side up on a flat surface. ① 2. Using a T-10 TORX driver, remove the four screws, (item ), that secure the receiver deck. ② 3. Pull the latch pin (item ) towards the opposite side of the analyzer to release the receiver deck.
  • Page 15 Step 6. Remove the Old Bottom Side Hardware and Cables • Remove the A17 LOMA 10 • Remove the A18 MA 20 • Remove the A22 Switch/Splitter • Remove the A23 and A24 Detectors • Remove the A27, A28, A29, and A30 First Converters (Mixers) •...
  • Page 16 Remove the A18 MA 20 ① 1. Using a 5/16-inch wrench, disconnect the two semirigid cables (item ) from the A18 MA 20. Remove these cables completely to allow the A18 MA 20 to be removed from the analyzer. Retain these cables for reinstallation later. ②...
  • Page 17 Remove the A22 Switch/Splitter ① 1. Disconnect the stranded switch control cable (item ) from the A22 switch/splitter. Remove this cable completely from the analyzer and discard it. It will be replaced later. ② 2. Using a 5/16-inch wrench, remove the five semirigid cables (item ).
  • Page 18 Remove the A23 and A24 Detectors ① 1. Disconnect the stranded control cable (item ) from each detector. Remove and discard this cable, it will not be used. ② 2. Using a 5/16-inch wrench, disconnect the remaining semirigid cable (item ) from each detector.
  • Page 19 Remove the A27, A28, A29, and A30 First Converters (Mixers) ① 1. Using a 5/16-inch wrench, disconnect the six semirigid cables (item ) and the four ② flexible cables (item ③ 2. Using a T-10 TORX driver, remove the four mounting screws (item ).
  • Page 20 Remove the A25 and A26 Test Port Couplers ① 1. Using a 5/16-inch wrench, remove the two semirigid cables (item ) from each of the test port couplers. Discard these cables, they will be replaced with new ones. ② 2. Using a 1-inch wrench, remove the test port coupler nuts (item ) from the front panel.
  • Page 21 Step 7. Install the New Hardware and Cables • Install the New A17 LOMA 12 • Install the New A18 MA 24 • Install the New A25 and A26 Test Port Couplers • Install the New A23 and A24 SOMA 70s •...
  • Page 22 Install the New A18 MA 24 1. Position the new A18 MA 24 in the analyzer as shown. Observe the below. CAUTION Be careful not to damage the center pins of the semirigid cables. Some flexing of CAUTION the cables is necessary when installing the assembly. Do not over-bend them. ③...
  • Page 23 Install the New A25 and A26 Test Port Couplers 1. Position the A25 and A26 test port couplers in the analyzer as shown. ① 2. Install the test port coupler nuts (item ) and, with a 1-inch torque wrench, tighten them to 72 in-lbs.
  • Page 24 Install the New A23 and A24 SOMA 70s ① 1. Place the SOMA 70 mounting brackets (item ) in the analyzer as shown. ② 2. Using a T-10 TORX driver, install three mounting screws (item ) in each bracket. Figure 16 SOMA 70 Brackets Installation Installation Note E8364-90022...
  • Page 25 3. Position the A23 and A24 SOMA 70s on the brackets as shown. ③ 4. Using a T-10 TORX driver, install two mounting screws (item ) in each SOMA 70. ➃ 5. Position a lid clamp (item ) on each SOMA 70. ➄...
  • Page 26 Install the New A22 Switch 1. Position the A22 switch in the analyzer as shown. ① 2. Using a T-10 TORX driver, install the four mounting screws (item ). Two screws were retained from removal of the A22 switch/splitter and an additional two are provided. 3.
  • Page 27 Install the New A27, A28, A29, and A30 First Converters (Mixers) 1. Remove all four mixers from the previously removed mixer bracket by removing three ① screws (item ) from each mixer. Retain the screws. ② 2. Remove the mixer bias board by removing the four attachment screws (item ).
  • Page 28 ④ 7. Connect the ribbon cable (item ) and place the mixer bracket, with all four mixers attached, into the analyzer. Observe the below. CAUTION Be careful not to damage the center pins of the semirigid cables. Some flexing of CAUTION the cables is necessary when installing the assembly.
  • Page 29 Install the New Cables Be very careful with these cables with 1.85 mm connectors. They are extremely CAUTION delicate and can be easily damaged. Using a 5/16-inch torque wrench set to 10 in-lbs, install the following new cables in the order listed.
  • Page 30 Figure 21 New Cables Installation Installation Note E8364-90022...
  • Page 31 Step 8. Lower the Receiver Deck ② 1. Pull the latch pin (item ) towards the opposite side of the analyzer to release the receiver deck. ② 2. Lift the receiver deck to partially lower it, then release the latch pin (item ).
  • Page 32 Step 9. Replace the Front Panel Nameplate and Overlay Remove the Old Nameplate ① 1. Using a sharp object such as a knife blade, lift one end of the old nameplate (item ). Be careful not to scratch the painted surface. 2.
  • Page 33 Remove the Old Overlay 1. From the back side of the front panel, use a blunt object in one of the cutouts in the lower ① frame to push the overlay (item ) and separate it from the front panel. 2.
  • Page 34 Step 10. Reinstall the Inner and Outer Covers Reinstall the Inner Cover 1. Place the analyzer top-side up on a flat surface. 2. Place the cover in position on the analyzer. ① 3. Using a T-10 TORX driver, install the eleven screws (item Figure 25 Inner Cover Reinstallation The figure above shows the E8364B front panel and floppy disk drive.
  • Page 35 Reinstall the Outer Cover This procedure is best performed with the analyzer resting on its front CAUTION handles in the vertical position. Do not place the analyzer on its front panel without the handles. This will damage the front panel assemblies. 1.
  • Page 36 Step 11. Enter the New Model Number Procedure Requirements • The analyzer must be powered up and operating to perform this procedure. • The Network Analyzer program must be running. • A keyboard and a mouse must be connected to the analyzer for this procedure. Enter the New Model Number 1.
  • Page 37 Step 12. Perform Post-Upgrade Adjustments The following adjustments must be made due to the hardware changes made to the analyzer. Perform these adjustments in the order listed. • 3.8 GHz PMYO frequency • LO power • offset LO • source calibration •...