Replacing The Solvent Line Filter - Thermo Scientific Vanquish VH-P10 Operating Manual

Table of Contents

Advertisement

7
Maintenance and Service

7.10.3 Replacing the Solvent Line Filter

When
Parts required
Preparations
Follow these steps
Page 168
Damage of solvent line filter
Impaired permeability of filter frit
As applicable:
Filter holder
Filter frit
Turn off the pump flow and wait until the system pressure is down to
zero.
1. Unscrew the cap of the solvent reservoir and remove the cap
together with the solvent line from the reservoir.
2. Remove the filter holder from the solvent line.
3. As applicable
Open the filter holder and remove the filter frit.
4. Assemble the solvent line filter, wearing appropriate clean gloves:
a) Place the frit in the filter holder (bottom part).
b) Make sure that the frit is in a level position.
c) Screw the filter top to the filter bottom.
Figure 67: Assembling the solvent line filter
No.
Description
1
Solvent line filter, bottom
2
Filter frit
3
Solvent line filter, top