Seal Wash System; Testing The Seal Wash System For Leakage - Thermo Scientific Vanquish VH-P10 Operating Manual

Table of Contents

Advertisement

7.5

Seal Wash System

7.5.1

Testing the Seal Wash System for Leakage

When
Items required
Follow these steps
Vanquish Pumps (VH-P10) Operating Manual
Follow the sequence of steps for the maintenance procedure that you
want to perform:
To test the seal wash system for leakage, see the next section
further down.
To replace the seal wash lines, see
To replace the seal wash detector (drop detector), see
page
129.
TIP To replace the seal wash seal, which are part of the pump head, see
section 7.6.7, page
146.
At regular intervals
If you performed maintenance on components in the pump head,
for example, after replacing the piston seals
Syringe (12 mL)
1. Disconnect the tubing from the seal wash detector.
Figure 35: Seal wash detector
7
Maintenance and Service
section 7.5.2, page
No.
Description
1
Seal wash detector
2
Tubing connection port
(port labeled in)
126.
section 7.5.3,
Page 125