7.3
General Rules for Maintenance and Service
Vanquish Pumps (VH-P10) Operating Manual
For successful maintenance and service procedures, follow these rules
and recommendations:
•
Before starting maintenance or service procedures, shut down the
pump when instructed to do so.
•
Use only the replacement parts specifically authorized and qualified
for the pump by Thermo Fisher Scientific. For ordering information,
see
section 10.4 Consumables and Replacement Parts, page
•
Follow all instructions step by step and use the tools recommended
for the procedure.
•
Before opening the flow path to replace capillaries in the system,
turn off the pump flow and wait until the system pressure is down to
zero.
•
Dirty components can contaminate the chromatography system.
Contamination leads to poor performance of the modules and entire
system or can even cause damage to the modules and system.
Therefore:
♦
Always wear appropriate gloves.
♦
Place the components only on a clean, lint-free surface.
♦
Keep your tools clean.
♦
Use only lint-free cloth for cleaning.
•
After you have performed maintenance or service on the flow
connections in the pump, test the pump for leakage before you
resume operation.
•
If you need to return the pump for depot repair, follow the
instructions in
section 7.15 Transporting or Shipping the Pump,
page
178.
7
•
Maintenance and Service
Page 117
211.