Updating The Pump Firmware - Thermo Scientific Vanquish VH-P10 Operating Manual

Table of Contents

Advertisement

7
Maintenance and Service

7.13 Updating the Pump Firmware

When
Items required
Preparations
Follow these steps
Page 174
Updating the pump firmware might be required, for example, when a
new firmware version is released that adds functionality or solves
problems of a previous version.
Firmware version/Chromeleon Service Release as appropriate
TIP When a new firmware version is released, the new version will be
included in the next available Chromeleon Service Release. The new
firmware will not be transferred automatically to the device when you
install the Chromeleon Service Release.
Read the release notes provided with the firmware and/or
Chromeleon Service release.
Verify the following:
The pump is connected in Chromeleon.
All operations on the instrument (Chromeleon 7) or timebase
(Chromeleon 6.8) that includes the pump have been stopped.
The instrument or timebase is idle.
The pump flow is turned off and the system pressure is down to
zero.
1. Start the Chromeleon 7 Instrument Configuration Manager or the
Chromeleon 6.8 Server Configuration program.
2. Perform a firmware update from the General tab page in the
configuration dialog box for the pump. For details, refer to the
Chromeleon Help.
Vanquish Pumps (VH-P10) Operating Manual