Performance Specifications - Thermo Scientific Vanquish VH-P10 Operating Manual

Table of Contents

Advertisement

9
Specifications
9.1

Performance Specifications

Type
Operating principle
Compressibility compensation
Flow range (settable)
Flow accuracy
Flow precision
Pressure range
Pulsation
Gradient formation
Proportioning
Accuracy
Precision
Number of solvents
Maximum stroke volume
Mixer volume
Dwell volume
Solvent degassing
Biocompatible
Communication
USB
I/O Interface
System Interlink
Control
Page 204
The pump performance is specified as follows:
Specification*
Parallel dual-piston pump with independent piston drives and variable
stroke volume
Fully automated, independent of the composition of the mobile phase
0.001 – 5 mL/min in 1 µL/min increments
± 0.1%
< 0.05% RSD or < 0.01 min SD, whichever is greater
5 – 151 MPa (50 – 1517 bar, 700 – 22002 psi)
< 0.4% or < 0.2 MPa, whichever is greater
High-pressure gradient proportioning
± 0.2% (of full-scale), for 0.2 to 4.0 mL/min from 1% to 99%
< 0.15%SD, for 0.2 to 4.0 mL/min from 1% to 99%
2 from 6
120 µL
Default configuration: 25 µL capillary mixer (proprietary)
35 µL
Internal 6-channel degasser
Yes
1 USB port (USB 2.0, "B" type connector)
1 USB hub with 3 ports (USB 2.0, "A" type connectors)
2 digital I/O ports (mini-DIN), each providing one input, one relay
output, and one bidirectional input/output
2 system interlink ports (RJ45-8 connectors)
Chromeleon 7, Chromeleon 6.8
The pump can be operated also with other data systems. For details,
contact the Thermo Fisher Scientific sales organization.
Keypad with 5 buttons for performing certain functions directly from
the pump
Vanquish Pumps (VH-P10) Operating Manual