5.6.4

Guiding Liquids to Waste

Vanquish Pumps (VH-P10) Operating Manual
The following liquids leave the pump through the outlets at the bottom
right of the leak tray:
Leaking liquid from the leak tray
Leaking liquid collects in the leak tray and flows off through the
drain port outlet into the drain system of the Vanquish system.
Waste liquid from the purge unit
Waste liquid from the purge unit flows off through the waste outlet
port into the drain system of the Vanquish system. To avoid false
leak alarms, verify that the waste line sits properly in the waste
outlet port. If it does not, liquid from the purge unit may accumulate
in the drain port and activate the leak sensor.
Figure 13: Outlets for guiding liquids to waste (view from top)
No.
Description
1
Waste line from purge unit: routing path to outlet
2
Drain port, with
2a
Drain port wall
2b
Groove in drain port wall for routing the waste line to the waste outlet
port
2c
Drain port outlet (to Vanquish drain system)
3
Waste outlet port (to Vanquish drain system)
4
Leak sensor
For information about how the liquid is discharged to waste through the
Vanquish drain system, refer to the Vanquish System Operating Manual.