Site Requirements

Power Considerations
Vanquish Pumps (VH-P10) Operating Manual
The operating environment is important to ensure optimal performance
of the pump. This section provides important requirements for the
installation site. Note the following:
Operate the pump only under appropriate laboratory conditions.
The pump is intended to be part of the Vanquish system. Observe
the site requirements for the Vanquish system as stated in the
Vanquish System Operating Manual.
For specifications, see the Specifications section in this operating
manual and the Operating Manuals for the other modules in the
Vanquish system.
The power supply of the device has wide-ranging capability, accepting
any line voltage in the range specified for the device.
CAUTION—Electric Shock or Damage to the Device
Connecting the device to a line voltage higher or lower than
specified could result in personal injury or damage to the device.
Therefore, connect the device to the specified line voltage only.
Never connect the device to a power socket that is shared with other
equipment (for example, multiple sockets).
Do not use extensions cords.
After the power to the device is turned off, the device is still
energized as long as the power cord is connected. Repair work on
the device while the device is connected to power could lead to
personal injury. Therefore, always unplug the power cord before
starting repair work inside the device. If you were instructed to
remove any covers or panels, do not connect the power cord to the
device while the cover or panels are removed.