Replacing The Seal Wash Detector - Thermo Scientific Vanquish VH-P10 Operating Manual

Table of Contents

Advertisement

7.5.3

Replacing the Seal Wash Detector

When
Parts required
Follow these steps
Vanquish Pumps (VH-P10) Operating Manual
4. Check the seal wash lines over the entire flow path:
a) Make sure that no bending, pinching or squeezing of the seal
wash lines is present at any point in the flow path.
b) If you have to cut tubing to length, use a tubing cutter. Make
sure that the cut is at right angle to the length of the line.
5. Flush the seal wash system (see
seal wash liquid.
Contamination of detector electrodes
Impaired functionality of the seal wash system
Seal wash detector assembly, consisting of the detector and funnel
Figure 37: Removing the seal wash detector assembly
1. Disconnect the tubing from the seal wash detector (port labeled in).
2. Unplug the detector cable from the Drop Det port.
3. Grasp and pull the seal wash detector assembly upward by the
funnel.
7
Maintenance and Service
section 5.6.6.3, page
No.
Description
1
Funnel, seal wash detector
2
Seal wash detector
3
Detector cable
4
Seal wash line
74). Use fresh