Installing The Slide-In Module - Thermo Scientific Vanquish VH-P10 Operating Manual

Table of Contents

Advertisement

7.16.2 Installing the Slide-In Module

Parts required
Tools required
Preparations
Vanquish Pumps (VH-P10) Operating Manual
NOTICE The packaging for the slide-in module differs from the
packaging for the complete module. Shipping the slide-in module or
the pump incorrectly leads to damage on the device. Always pack
and ship the slide-in module and the pump in the respective,
dedicated packaging.
CAUTION—Possible Contamination
Hazardous substances may have contaminated the device during
operation and may cause personal injury to service personnel.
Decontaminate all parts of the device that you want to return for
repair.
Fill in and sign the Decontamination Certificate, which is part of the
Service Return Form. Sign the certificate to confirm that the device
has been properly decontaminated and that it is free of hazardous
substances.
Thermo Fisher Scientific refuses to accept devices for repair if the
Decontamination Certificate is missing.
Replacement slide-in module
Screwdriver, Torx T20
1. Verify that the pump enclosure into which the slide-in module is
installed is clean. If required, clean the inner and outer surfaces of
the enclosure. See the Cleaning the Pump section in this operating
manual.
2. When installing the slide-in module to an enclosure in the system
stack, check that the enclosure is placed correctly in the stack.
7
Maintenance and Service