Inline Filter; Testing The Inline Filter For Permeability - Thermo Scientific Vanquish VH-P10 Operating Manual

Table of Contents

Advertisement

7.8

Inline Filter

7.8.1

Testing the Inline Filter for Permeability

When
Additional items required
Follow these steps
Vanquish Pumps (VH-P10) Operating Manual
Follow the sequence of steps for the maintenance procedure that you
want to perform:
To test the inline filter for permeability, see the next section.
To replace the inline filter or the capillary mixer (volume: 25 µL)
connecting the purge unit to the filter inlet, see
page
160.
If you observe pressure pulsation, inconstant pressure, or high
backpressure at the column and pump
Water as solvent
Tissue
1. Disconnect the capillary on the outlet of the inline filter.
2. To absorb the liquid leaving the filter, place the tissue under the
filter outlet.
3. Apply a flow of 2 mL/min using water.
4. Observe the pump pressure.
5. Depends on the pressure reading:
Pressure is less than 0.8 MPa
Continue with the next step.
Pressure is 0.8 MPa or higher
Replace the inline filter.
6. Reconnect the capillary to the outlet of the inline filter.
7. Test the pump for leakage (see
7
Maintenance and Service
section 7.8.2,
section 7.11, page
170).
Page 159