5.2

Installing the Pump

Vanquish Pumps (VH-P10) Operating Manual
The Vanquish system is installed and set up by a Thermo Fisher Scientific
service engineer, including all modules and options or parts shipped with
them. The service engineer checks that the installation is correct and
that the Vanquish system and modules operate as specified. The
engineer also demonstrates the basic operation and main features.
If personnel other than a Thermo Fisher Scientific service engineer
installs the pump, follow the steps below.
NOTICE The pump is part of the Vanquish system. Therefore, follow the
order for installing the system modules as described in the Vanquish
System Operating Manual.
1. Pay attention to the safety guidelines and observe all site
requirements.
For the safety guidelines when installing the pump, see
page
48. For the site requirements, see
2. Set up the pump hardware. See
3. Set up the flow connections. See
4. Turn on the pump. See
TIP Before turning on the power to a Vanquish system module for the
first time, verify that the chromatography software is installed on the
data system computer. When the power is turned on, the required USB
drivers are automatically found and the Windows® operating system can
detect the device.
5. Set up the pump in the software. See
6. Recommended:
Perform Instrument Installation Qualification.
In Chromeleon, a wizard is available to guide you through the
qualification process:
On the Chromeleon 7 Console: Click Tools > Instrument
Qualification > Installation Qualification.
section 5.3, page
section 5.5, page
section 5.6, page
section 5.7, page
77.
section 5.8, page
5
Installation
section 5.1,
51.
54.
58.
78.
Page 49