Resolving Leakage - Thermo Scientific Vanquish VH-P10 Operating Manual

Table of Contents

Advertisement

8.3

Resolving Leakage

When
Preparations
Parts required
Additional items required
Follow these steps
Vanquish Pumps (VH-P10) Operating Manual
The leak sensor is wet. The leak sensor reports leakage.
If leakage is not resolved after 3 minutes, the pump stops the flow.
When resolving leakage, observe the safety guidelines and general rules
for maintenance and service as presented in
Service, page
113.
As leakage usually occurs at a connection, inspect all components and
connections in the flow path.
Replacement part as required
Cloth or tissue
1. Locate the source of the leak.
The following tests can help you to identify the source:
Test the seal wash system for leakage (see
page
125).
Test the pump for piston seal leakage (see
page
153).
Test the entire pump for leakage (see
2. Tighten or replace the connection or component as required.
3. With a cloth or tissue, thoroughly absorb all liquid that has collected
in the leak tray and under the leak sensor. Be careful not to bend the
sensor.
4. Allow the sensor to adjust to the ambient temperature for a few
minutes.
5. If leakage is no longer reported, you can resume operation.
8
Troubleshooting
chapter 7 Maintenance and
section 7.5.1,
section 7.6.9,
section 7.11, page
170).
Page 197