Installation
5.6.5

Connecting the Solvent Lines

Parts and tools required
Preparation
Page 64
The pump is shipped from the manufacturing site with the solvent lines
installed between the components in the pump.
To complete the setup, install the solvent lines between the degasser
inlet ports and the solvent reservoirs.
Solvent reservoir assemblies, including reservoir, reservoir cap, cap
plugs, and retaining guide for the solvent line
Solvent lines
Solvent line filters
Optional to shut off the solvent flow through the system at the
solvent reservoir
Shut off valve, kit containing one shut off valve and fitting
connections
Tubing cutter
If you want to install shut off valves after initial setup or operation of
the pump, follow the steps in
Off Valves, page
106.
TIP To avoid the formation of gas bubbles in the solvent reservoirs, place
the reservoirs in the system stack on the same level or higher than the
pump. The setup description below assumes that the reservoirs are
placed in the solvent rack.
Assemble the solvent line filter, wearing appropriate clean gloves:
a) Place the frit in the filter holder (bottom part).
b) Make sure that the frit is in a level position.
section 6.9.3 Installing Solvent Shut
Vanquish Pumps (VH-P10) Operating Manual