Operating Instructions; Preparation; Connection - MAGSTIM MOP11-EN Operating Manual

Alpha coil range
Table of Contents

Advertisement

SECTION 4
Magstim Alpha Coil Range
Operating Manual
OPERATING INSTRUCTIONS
Note: For more detailed operating instructions see the Operating Manual for the appropriate
stimulator.
Note: to avoid problems with interference, the stimulating coils should not be used in the vicinity of
any equipment not in compliance with the EMC Standard EN 60601-1-2, including mobile phones.
Note: Please ensure you are familiar with all sections of this operating manual prior to following these
operating instructions.
4.1
Preparation
At the start of each session the operator must check the external casing surrounding the coil for any
signs of damage. If any cracks are visible in the cover or resin coating, or there is damage to _the
handle or cable, the item must not be used and should be returned to The MagstIm Company Limited
for servicing and repair.
Note: The contacts in the coil connector, and the stimulator coil socket should be checked before
connecting to the system for any signs of tarnishing or burning. See Section 5.2 for an example of
contact burning.
4.2
Connection
1.
Place the coil into the coil holder situated on the right hand side of the stimulator as shown in
Figure 4.1. (Coil holder is situated on the right hand side of the master unit on the Bistim
2
)
~,
-
Figure 4.1: Correct method of placing coil in holder.
2.
Connect the coil to the stimulator (Bistim
2
connecting module when using the Magstim Bistim
2
)
by
aligning the locating key on the coil connector with the corresponding receptor_ on _the coil output
socket. It is then possible to rotate the outer collar of the coil connector clockwise
in
order to lock
the coil to the stimulator.
Coil locating key
Coil output
socket receptor
Figure 4.2: Locating key on the coil connector (left} and the stimulator coil socket (right)
© The Magstim Company Limited
6
MOP11-EN-01