Operating Instructions; Preparation; Connection - MAGSTIM 2002 Operating Manual

Hide thumbs Also See for 2002:
Table of Contents

Advertisement

SECTION 4 OPERATING INSTRUCTIONS
Magstim 200
2
Operating Manual
Note: to avoid problems with interference, the stimulating coils should not be used in the vicinity of
any equipment not in compliance with the EMC Standard EN 60601-1-2, including mobile phones.
Note: Please ensure you are familiar with all sections of this operating manual prior to following these
operating instructions.
4.1
Preparation
1.
At the start of each session the operator must check the Magstim 200
2
and any accessories used
for any signs of external damage and to identify any cleaning required. If any cracks are visible in
the housing, or there is damage to any of the cables, the items must not be used and should be
returned to The Magstim Company Limited for servicing and repair. If cleaning is required, please
follow the instructions in Section 5 of this Operating Manual.
2.
Ensure that the Magstim 200
2
unit is switched off on the rear.
4.2
Connection
1.
If a foot switch is to be used, insert the pneumatic connector of the foot switch to the rear panel
Foot Switch Socket on the Magstim 200
2
unit.
2.
If an external triggering or recording device is to be used, connect it to the Isolated Interface on
the rear of the Magstim 200
2
unit. Refer to Appendix B for more details on the Isolated Interface.
3.
Connect the stimulating coil to be used to the Magstim 200
2
unit. There is locating key on the coil
connector, which needs to be aligned with the corresponding receptor on the stimulator coil
socket. It is then possible to rotate the outer collar of the coil connector in order to lock the coil to
the stimulator.
Coil
Stimulator socket receptor
Figure 4.1: Locating key on the coil connector (left) and the stimulator coil socket (right)
Care must be taken to attach the coil connector to the stimulator correctly (see Figure 4.2).
Failure to do so can cause damage to the connector pins and in severe cases, internal damage to
the coil and the stimulator.
Figure 4.2: Diagram showing correct and incorrect ways to connect the coil to the stimulator
4.
Connect the power lead to the rear of the Magstim 200
2
unit.
© The Magstim Company Limited