Semicritical medical devices

Manual cleaning
Perform manual cleaning preferably under running
water and with commercially available cleaning
agents (pH value ≤12).
Procedure:
1 Wash off surface dirt under running water.
2 Use cleaning agents in accordance with the
manufacturer's instructions. Make sure that all
surfaces and interior spaces which must be
cleaned are reached. If necessary, use suitable
brushes.
3 Rinse items thoroughly under running water un-
til cleaning agent residues are no longer dis-
cernible.
4 Check parts for visible dirt and damage. If nec-
essary, repeat manual cleaning.
Manual disinfection
For choosing the appropriate disinfectant, observe
country-specific lists of disinfectants. The list of the
German Association for Applied Hygiene (Verbund
für Angewandte Hygiene VAH) applies in German-
speaking countries.
Strictly observe the manufacturer's instructions for
using disinfectants. The composition of disinfect-
ants may change.
Procedure:
1 Immerse items in disinfectant.
2 After the contact time has elapsed, rinse items
thoroughly under running water until disinfect-
ant residues are no longer discernible.
3 Check parts for visible dirt and damage. If nec-
essary, repeat manual disinfection.
4 Thoroughly shake out residual water. Allow
items to dry thoroughly.
Instructions for use Primus Infinity Empowered SW 4.5n
Cleaning, disinfection and sterilization
Machine cleaning and disinfection
Perform machine cleaning and disinfection using a
washer-disinfector in accordance with
EN ISO 15883, preferably with a cart for anesthe-
sia accessories and ventilation accessories.
Procedure:
1 Observe instructions for use of the washer-dis-
infector.
2 Securely position items in the basket. Make
sure that all interior spaces and surfaces are
completely flushed and that water can drain off
freely.
3 Use suitable cleaning agent.
4 Select suitable program, preferably anesthesia
program.
– Cleaning must be carried out at 40 °C to
60 °C (104 °F to 140 °F) for at least 5 min.
– Thermal disinfection must be carried out at
80 °C to 95 °C (176 °F to 203 °F) and with
corresponding contact time.
5 Carry out final rinsing with deionized water.
6 Immediately remove items from the washer-dis-
infector.
7 Check parts for visible dirt and damage. If nec-
essary, repeat program or perform manual
cleaning and disinfection.
8 Allow items to dry thoroughly.
WARNING
Risk of device failure and patient injury
Correct operation of the workstation may be
impaired and lead to failure of the workstation
if the control areas in the valve plate are not
dried completely.
The valve plate must be sterilized after wash-
ing in order to dry it.