Dräger Primus Infinity Empowered Instructions For Use Manual page 133

Anesthesia workstation
Hide thumbs Also See for Primus Infinity Empowered:
Table of Contents

Advertisement

Mode
Volume, Volume AF,
Alarm
Press. Support
inDes
[Vol%]
inSev
[Vol%]
P
AW
[hPa
(cmH
O)]
2
APNEA PRES-
SURE
APNEA FLOW
APNEA CO
2
1) In Standby, these alarms can be configured ON or OFF for switching to Man. Spont., Ext. Outlet, and Monitoring. When
the alarm limits are set to ON the value is adopted from the automatic ventilation mode. The default value for this config-
uration is OFF.
2) In Man.Spont., Monitoring, and Ext. Outlet, the alarm is triggered after 65 seconds.
– – The factory setting is outside the monitored range; the corresponding alarm limit is disabled.
All apnea, apnea pressure, apnea flow, and apnea
CO
alarms are active after at least 15/30 seconds
2
depending on the set frequency/Freq
All apnea and limit-based O
alarms are only active if a breath has already been
detected (AutoWakeUp).
Instructions for use Primus Infinity Empowered SW 4.5n
Press.
Pressure,
Support
CPAP
ON
ON
ON
ON
ON
ON
ON
OFF
ON
ON
.
MIN
, CO
, N
O, and agent
2
2
2
Ext. Outlet
ON
ON
1)
ON
ON
ON
1)
ON
ON
ON
ON
OFF
OFF
ON
OFF
2)
ON
ON
Monitoring,
Factory setting
Man.Spont.
ON
12.0
1)
– –
ON
4.2
1)
– –
ON
ON
OFF
OFF
- - -
2)
ON
- - -
Alarms
40
8
8
133