Connecting The Endotracheal Aspiration System (Optional); System (Optional) - Dräger Primus Infinity Empowered Instructions For Use Manual

Anesthesia workstation
Hide thumbs Also See for Primus Infinity Empowered:
Table of Contents

Advertisement

Assembly and preparation
WARNING
Risk of patient injury
If the side openings of the receiving system
are blocked, negative pressure may result in
the breathing system and the patient's lungs.
Always make sure the side openings of the re-
ceiving system are not blocked.

Connecting the endotracheal aspiration system (optional)

Prepare the endotracheal aspiration system ac-
cording to the instructions for use included with
the system.
Depending on the aspiration version used:
A
If Air is used as driving gas:
Secure the Air connecting hose of the endotra-
cheal aspiration system to the central gas sup-
ply (CS) for Air.
62
AIR
The anesthetic gas scavenging system may be
connected optionally to the left-hand side of the
workstation.
NOTE
If the anesthesia machine is operating on an in-
cline exceeding 5°, make certain the AGS is func-
tioning correctly, e.g., by means of visual inspec-
tion.
Optionally, the Air connecting hose of the en-
dotracheal aspiration system can be connected
to the AIR outlet (A) at the back of the Primus
Infinity Empowered.
For vacuum-driven aspiration:
Connect the vacuum hose of the endotracheal
aspiration system directly to the terminal unit.
WARNING
Risk of patient injury
If not used correctly, the suction unit may in-
jure the patient.
Prior to use, disconnect the patient from the
ventilator, and pay special attention to the in-
structions for use of the suction unit.
Instructions for use Primus Infinity Empowered SW 4.5n