Dräger Primus Infinity Empowered Instructions For Use Manual page 181

Anesthesia workstation
Hide thumbs Also See for Primus Infinity Empowered:
Table of Contents

Advertisement

E pulse volume (optional)
0 = off
to
9 = maximum volume
F alarms on/off
CO
and SpO
alarms (optional) and HLM
2
2
mode, see page 135, can be enabled and disa-
bled.
The flow apnea alarm can be configured to be a
low-priority alarm (technical alarm). This setting
remains active until switching to standby and
starting a new case.
This configuration is only possible when the set-
ting immediate alarm upon apnea? (patients
<6 years) is set no (see page 174).
The paramet. settings (A) menu contains the fol-
lowing sub-menus:
Volume
B
B Scaling amplitude
– CO
, P
, flow, O
2
AW
The setting is made automatically or by se-
lecting a pre-set scale.
auto: Automatic adjustment to the next
higher or lower scale after two passes if the
scaling frame is exceeded.
Instructions for use Primus Infinity Empowered SW 4.5n
C
D
A
2
– tidal volume
auto: A suitable scaling is selected auto-
matically in accordance with the set age.
<1 year: 50 mL,
1 to 2 years: 150 mL,
>2 to 10 years: 500 mL,
>10 years: 1000 mL.
C Units
– CO
: mmHg, Vol%, kPa
2
– P
: hPa, mbar, cmH
AW
– agents: Vol%, kPa
D agent monitoring
– MAC display
(See page 147 for a detailed description of
the MAC definition and calculation.)
– Related to age
The logbook entries menu (E) contains the follow-
ing sub-menus:
Volume
F
F logbook entries triggered by
– Time interval (min)
Entries are triggered when a fixed time inter-
val expires (in minutes).
– Alarm
Entries are triggered when a warning is is-
sued.
– Caution
Entries are made when a caution message
is issued.
Configuration
O
2
E
G
181