Dräger Primus Infinity Empowered Instructions For Use Manual page 204

Anesthesia workstation
Hide thumbs Also See for Primus Infinity Empowered:
Table of Contents

Advertisement

Remedying faults
Priority
Message
Caution
INSP. HAL. LOW
(15)
INSP. ISO. LOW
INSP. ENF. LOW
INSP. DES. LOW
INSP. SEV. LOW
Caution
INSP. N
O HIGH
2
(12)
Caution
INSP. O
2
(12)
Warning
INSP. O
2
(31)
Warning/
INTERNAL TEMP. HIGH *
Note
(29/10)
Caution/
LOSS OF DATA *
(14/–)
204
HIGH
LOW
Cause
The inspiratory anesthetic
gas concentration has
fallen below the lower
alarm limit for at least
2 breaths.
Inspiratory N
O concentra-
2
tion exceeds the upper
alarm limit of 82 %.
Inspiratory O
concentra-
2
tion exceeds the upper
alarm limit.
Inspiratory O
concentra-
2
tion is below the lower
alarm limit.
Temperature inside the de-
vice is too high.
Fan is defective.
Extreme, non-physiological
ventilation settings.
Data loss of the settings
and/or the system configu-
ration.
Instructions for use Primus Infinity Empowered SW 4.5n
Remedy
Check the vaporizer and
fresh-gas settings.
Check for leakages in
breathing system.
Check soda lime (dried
out?).
Check N
O concentration
2
in the fresh-gas flow.
Flush.
Check O
concentration in
2
the fresh-gas flow.
Check O
concentration
2
and fresh-gas setting.
Check for leakages in
breathing system.
Check O
supply.
2
Check ambient conditions.
Ensure air circulation at
back of device.
Call DrägerService.
Check ventilation setting.
Check the current settings
and the default settings.
Repeat settings if neces-
sary.
Call DrägerService.
Alarm can be canceled by
pressing
.