Dräger Primus Infinity Empowered Instructions For Use Manual page 257

Anesthesia workstation
Hide thumbs Also See for Primus Infinity Empowered:
Table of Contents

Advertisement

Delta V
T
Measuring range
Resolution of the displayed value
Accuracy
Volume VT
INSP
Measuring range
Resolution of the displayed value
Accuracy
Minute volume MV
Measuring range
Resolution of the displayed value
Accuracy
Compliance C
PAT
Measuring range
Resolution of the displayed value
Accuracy
MV
LEAK
Measuring range
Resolution of the displayed value
Accuracy
MV
MAND
Measuring range
Resolution of the displayed value
Accuracy
Instructions for use Primus Infinity Empowered SW 4.5n
0 to 9999 mL
1 mL
±16 % or ±10 mL, the greater value applies
0 to 9999 mL
1 mL
±8 % of measured value or ±5 mL, the greater value applies
0 to 99.9 L per minute
0.1 L per minute
±8 % of the measured value or ±0.1 L per minute, the greater
value applies
0 to 250 mL/hPa (0 to 250 mL/cmH
0.1 mL/hPa (0.1 mL/cmH
±15 % of the measured value or ±0.5 mL/hPa
(±0.5 mL/cmH
O), the greater value applies
2
0 to 9.99 L per minute
0.01 L per minute
±15 % of (MV
+ MV
EXP
greater value applies
0 to 99.9 L per minute
0.1 L per minute
±8 % of measured value or ±0.1 L per minute, the greater
value applies
O)
2
O)
2
) or ±0.1 L per minute, the
LEAK
Technical data
257

Advertisement

Table of Contents
loading

This manual is also suitable for:

Sw 4.5n