Ventilator and gas delivery

The following parameters can be set in the menu
ventilator and gas delivery > ventilator and gas
delivery:
A parameter default values
Standby Conf.
A
B
C
D
E
F
G
– Volume Mode (B)
– Volume AF Mode (C)
– Pressure Mode (D)
– Pressure Support (E)
– Patient attributes (F)
Select mode via rotary knob and confirm. Soft
keys for ventilation parameters appear. Set ven-
tilation parameters, see page 38.
The factory settings for ventilation parameters can
be found in "Operation".
The trigger sensitivity can be set separately in the
available ventilation modes.
If the trigger has been pre-set to OFF in Volume
Mode, Volume AF Mode or Pressure Mode, the
value configured under pressure support will auto-
matically be adopted when synchronization is acti-
vated during operation. The same also applies with
regard to adopting the value for
cannot be configured in the Volume Mode, Vol-
ume AF Mode and Pressure Mode.
– Gas delivery (G)
Instructions for use Primus Infinity Empowered SW 4.5n
H gas supply checks
Which gas supplies have been connected can be
determined in this menu:
P
although this
PS
Select and confirm via the rotary knob.
The soft keys for O
The carrier gas is selected via the Air or
N
O keys and confirmed via the rotary knob.
2
Standby Conf.
H
I
J
K
L
M
N
Connected gas supplies
I
O
line: yes/no
2
J Air line: yes/no
K N
O line: yes/no
2
L O
cylind.: yes/no
2
M Air cylind.: yes/no
N N
O cylind.: yes/no
2
NOTE
Only the gas supply defined as being present in
the configuration will be included in the self test.
Configuration
and Flow appear.
2
O
Factory
setting
yes
yes
yes
yes
no
no
175