Dräger Primus Infinity Empowered Instructions For Use Manual page 207

Anesthesia workstation
Hide thumbs Also See for Primus Infinity Empowered:
Table of Contents

Advertisement

Priority
Message
Warning/
NO AIR DELIVERY *
Note
(25/10)
Warning/
NO N
O DELIVERY*
2
Note
(25/10)
Warning
NO O
DELIVERY
2
(31)
Warning
NO SpO
2
(31)
Instructions for use Primus Infinity Empowered SW 4.5n
Cause
Compressed air supply has
failed.
The central gas supply is
not connected or the com-
pressed air hose is kinked.
Optional air cylinder is
empty or closed.
Compressed air compres-
sor has failed.
N
Probe for the central gas
supply not plugged in or
N
N
closed.
O
Probe for the central gas
supply not plugged in or O
hose kinked.
O
closed.
PULSE
No pulse signal detected
with the SpO
ment for approx.
10 seconds.
NiBP measurement on the
same arm.
O supply has failed.
2
O hose kinked.
2
O cylinder empty or
2
supply has failed.
2
cylinder empty or
2
measure-
2
Remedying faults
Remedy
Open optional Air reserve
gas cylinder.
Check central gas supply,
see page 187.
Check connection to cen-
tral gas supply.
Connect a full air cylinder
or open the cylinder valve.
Check compressor.
Open N
O reserve gas cyl-
2
inder.
Check central gas supply,
see page 187.
Check connection to cen-
tral gas supply.
Connect a full N
O cylinder
2
or open the cylinder valve.
Open O
reserve gas cylin-
2
der.
Check central gas supply,
see page 187.
Check connection to cen-
tral gas supply.
2
Connect a full O
cylinder
2
or open the cylinder valve.
Check patient's condition!
Check application of SpO
sensor.
Measure blood pressure on
the other arm.
2
207