Supplement to the instructions for use
Atlan
A100 (XL), A300 (XL), A350 (XL)
Instructions for use
Atlan A300, A300 XL, A350,
A350 XL SW 1.0n
Atlan A300, A300 XL, A350,
A350 XL SW 2.0n
Atlan A100 (XL), A300 (XL),
A350 (XL) SW 2.1n
 Keep this supplement with the instructions for use.
This supplement updates the information of the instructions for use in the following
chapters.

Assembly and preparation

Safety instructions
Electrical safety
The following section has been changed:
Battery supply
If the battery is not sufficiently charged, it may not be possible to maintain operation
for the required period if the mains power supply fails. As a result, the patient could
be put at risk.
► Before first operation or after storage, charge the battery for at least 8 hours.
► Avoid short charge and discharge cycles.
► When the device is switched on, do not disconnect it from the mains power
supply.
► When the device is powered by the internal battery while switched on, ensure
the functional integrity of the battery by performing a battery test at regular
intervals. For further information see: "Inspection", page 6.
|
Supplement
Atlan A100 (XL), A300 (XL), A350 (XL)
Supplement to the instructions for use
Part number
Edition
9056001
1 – 2019-02 and higher
9510601
1 – 2021-06 and higher
9511481
1 – 2022-08 and higher
|
English
5