Overview
WARNING
Risk of electric shock
Connecting devices to the auxiliary outlets of
the anesthesia machine can cause an in-
crease in leakage current beyond permissible
values if the protective conductor of a device
fails.
Check the leakage current when connecting
devices to the auxiliary outlets. If connecting
a device (or devices) increases the leakage
current to a value which exceeds the permis-
sible value, do not use the auxiliary outlets of
the anesthesia machine; use a separate wall
socket instead.
The system must fulfill the requirements for
medical electrical equipment in accordance
with the relevant standards, see chapter
"Technical data", page 270.
29
Instructions for use Primus Infinity Empowered SW 4.5n