Remedying faults
Priority
Message
Caution/
INCORRECT FG FLOW *
Note
(14/10)
Caution
INSP. CO
(11)
Note
INSP. FLOW SENSOR FAIL
(8)
202
HIGH
2
Cause
Set fresh-gas flow cannot
be delivered.
Soda lime in circle system
exhausted.
Leak or fault in breathing
system.
High respiratory rates.
If the respiratory rates are
high, the measured value
can no longer follow the
gas concentration properly
due to the design of the
system.
Dead space ventilation.
Insp. flow sensor is defec-
tive.
Instructions for use Primus Infinity Empowered SW 4.5n
Remedy
Check vaporizer setting.
Make sure the emergency
O
delivery is closed.
2
Set fresh-gas flow to be-
tween 3 and 10 L/min.
Increase the fresh-gas flow.
Replace soda lime.
Replace breathing system!
Adjust alarm limits if neces-
sary.
Check ventilation settings.
Replace flow sensor,
see page 220.