Dräger Primus Infinity Empowered Instructions For Use Manual page 93

Anesthesia workstation
Hide thumbs Also See for Primus Infinity Empowered:
Table of Contents

Advertisement

Data received from breathing circuit
Mandatory ventilation
Synchronized Intermittend Man-
datory Ventilation
Assisted ventilation
Mandatory Minute Volume Venti-
lation
Continuous positive airway pressure
Continuous positive airway pressure
(except proportional pressure support)
Independent Lung Ventilation
1) Nomenclature may vary dependent on device.
2) If a ventilation mode software option has not been activated, a substitute mode will be selected.
If Pressure Support (exception: Proportional Pres-
sure Support) has been activated in order to sup-
port the patient's spontaneous breathing efforts,
Primus Infinity Empowered will also take this set-
ting, as long as the software option has been acti-
vated. Volume Mode AutoFlow (Volume AF) will be
transferred into Volume Mode, if the software op-
tion is not activated.
Fresh-gas settings
Data for fresh-gas settings from devices without re-
breathing systems, will not be transferred to the
Primus Infinity Empowered.
Instructions for use Primus Infinity Empowered SW 4.5n
1)
Ventilation mode
Volume-controlled ventilation
Volume guaranteed
Pressure controlled
Volume-controlled ventilation
Volume guaranteed
Pressure controlled
with Pressure Support
Continuous Flow
External Fresh Gas
Man./Spont.
Resulting mode Primus Infinity
otherwise Volume Mode
otherwise Volume Mode,
Press. Support
otherwise Man.Spont. mode
Press. Support
otherwise Pressure Mode
otherwise Man.Spont. mode
otherwise Man.Spont. mode
Alarm limits and alarm activation
The numerical values are taken as is, for the follow-
ing alarm limits:
– minute volume high and low
– airway pressure high and low
– inspiratory O
concentration high and inspira-
2
tory O
concentration low
2
Whether or not these alarm limits will be valid, de-
pends on how Primus Infinity Empowered has been
configured.
Operation
Empowered
Volume Mode
2)
Volume AF
,
Pressure Mode
Volume Mode,
Trigger ON
2)
Volume AF
,
Trigger ON
Pressure Mode,
Trigger ON
2)
,
CPAP
2)
,
CPAP
ignored completely
2)
Ext. Outlet mode
,
2)
Ext. Outlet mode
,
Man.Spont. mode
93