Basic settings and audible signals

The menu basic settings audible signals con-
tains the following submenus:
Standby Conf.
A
B C
A alarm volume
The minimum alarm volume can be set to a
value between 1 to 9. The standard alarm vol-
ume cannot be set below this limit.
Alarm volume
Minimum: 1 to 9
Standard: 1 to 9
The standard volume cannot be set lower than
the minimum volume.
CAUTION
Risk of use error
If using features like the 'breathing sound' or when
operating under loud ambient conditions, the au-
ditory alarm signals may not be heard.
Always set the volume of the alarm signal suffi-
ciently high.
The alarm NO O
DELIVERY is always announced
2
at the maximum volume.
Primus Infinity Empowered takes into account the
national regulations of certain countries which re-
quire a minimum volume of 45 dB(A). Settings 1 to
3 are not available for these countries.
Instructions for use Primus Infinity Empowered SW 4.5n
D
Factory setting
4
5
However, the minimum alarm volume can be set to
values between 1 to 3 by service personnel, if re-
quired.
B breathing sound (optional)*
0 = off
to
9 = maximum volume
Factory setting 0
The breathing sound is generated by a breathing
sound module. This module converts the measured
inspiratory and expiratory flow values into audible
sounds similar to the sound of breathing.
The volume depends on the set patient age. This
dependency allows optimal volume at all flow lev-
els.
All alarms are still audible with the breathing sound
volume set to the maximum.
C pulse volume (optional)
0 = off
to
9 = maximum volume
Factory setting 0
D date/time language
– date/time
month, year
hours : minutes
– time format
Factory setting 24 h
– language
of the display texts
Factory setting English (GB)
*
Only in connection with breathing sound module.