Alarms

Alarm displays

A
Alarm messages are displayed in the alarm field (A)
in order of priority.
All displayed alarms are sorted according to the
three classes defined on page 129. Within these
classes, the alarms are sorted and displayed ac-
cording to an internal priority system. A priority of
31 indicates the highest, a priority of 1 the lowest
priority. The priority numbers are given in the table
"Alarm – Cause – Remedy" on page 194. The inter-
nal priorities are not displayed.
Up to three messages can be displayed simultane-
ously. In some cases, the corresponding measured
values are highlighted on the screen by a flashing
background in addition to the alarm message.
If more than three alarms occur simultaneously, the
more
symbol
appears (B) to the right of the alarm
field and the soft key show all alarms (C) is acti-
vated on the right-hand side of the screen.
When this soft key is pressed, the upper curve dis-
play (D) is replaced by up to six additional alarm
fields for 15 seconds.
128
B
C
If more than nine alarm messages are simultane-
ously active, the lowest priority alarms will not be
displayed until the total number of active alarm
messages falls below nine.
The alarm tone sequence accompanying a dis-
played alarm message with the highest priority will
always be sounded at least once completely. The
alarm tone sequences of alarm messages with
lower priorities will not sound if a higher priority
alarm is activated, i.e. the tone sequence thereof
will sound.
If an alarm message of the same class as an active
alarm message is generated, the alarm tone of the
new alarm only sounds if the priority is higher than
the priority of the previously active alarm.
Instructions for use Primus Infinity Empowered SW 4.5n