Inspection

Perform inspections at regular intervals and ob-
serve the following specifications.
Checks
Inspection and safety checks
1) Designation applies to the Federal Republic of Germany; corresponds to the "Recurring safety inspection" in the Federal
Republic of Austria

Safety checks

Safety checks are no substitute for preventive
maintenance measures (including preventive re-
placement of wear parts) as identified by the man-
ufacturer.
WARNING
Risk of medical device failure
If safety checks are not performed on a regu-
lar basis, the proper operation of the medical
device can be compromised.
Perform safety checks at the indicated inter-
vals.
1 Check accompanying documents:
– Instructions for use are available
2 Perform a functional test of the following fea-
tures according to the instructions for use:
– Perform self test.
– Check O
measurement.
2
– Check CO
measurement and AGAS meas-
2
urement.
– Check flow measurement.
– Check pressure measurement.
– Check battery backup for function.
– Check concentration delivery of anesthetic
vaporizer.
Instructions for use Primus Infinity Empowered SW 4.5n
Interval
1)
Every 12 months
Personnel responsible
Experts
– Check interlock function on vaporizer plug-
in system.
3 Check that the device combination is in good
condition:
– All labels are complete and legible
– There is no visible damage
– Fuses which are accessible from the out-
side are in compliance with the specified
values
4 Using the instructions for use, check that all
components and accessories needed to use the
product are available.
5 Check the electrical safety according to
IEC 62353.
6 Check safety features:
– Correct functioning of the alarm generator
Maintenance