The following parameters can be set in the menu
ventilator and gas delivery > weight related set-
tings > body weight related ventilator settings:
Standby Conf.
E
F
G
H
E body weight related ventilator settings
– V
and freq. presetting related to ideal
T
body weight: (F)
If the settings for V
ferred to the patient's body weight, the initial
value for V
can be selected in accordance
T
with the Radford nomogram.
Factory setting: yes
– preset configuration: (G)
Instructions for use Primus Infinity Empowered SW 4.5n
I
and freq. are to be re-
T
Select, edit, and confirm the V
via the rotary knob.
The settings for V
T
between the four predetermined classes.
weight
(ideal BW)
[kg]
Setting
ranges
2
10 to 25
15
60 to 150
65
300 to 500
100
550 to 800
– restore factory default presets (H).
Select and confirm to restore the factory set-
tings.
The default settings are activated immediately
upon exiting the configuration menu.
Press exit (I), to exit the menu.
Configuration
to be changed
T
are interpolated for weights
V
[mL]
Freq.
T
Factory
[1/min]
settings
15
35
110
26
450
13
700
10