1 Press the button (A); the mounting swings
open.
2 Before fitting, shake the disposable absorber,
e.g., by turning it upside down several times in
order to loosen up the soda lime.
3 Remove the seal from the new disposable ab-
sorber.
4 Slide the new disposable absorber into the
mount (B) and
5 Push the absorber into the machine until it en-
gages.
If the Infinity ID CLIC absorber has been configured
in Standard Conf. (see "Soda lime depletion"
on page 80), the new absorber will be automatically
detected by Primus Infinity Empowered.*
If the Infinity ID CLIC absorber was not configured
6 Reset the soda lime change log to current date
by pressing the soft key soda lime changed,
see page 88.
*
Function not available at present.
Instructions for use Primus Infinity Empowered SW 4.5n
Reusable absorber
A
B
7 Push the insert fully into the absorber (C).
8 Fill the absorber to the upper mark with fresh
Recommendation:
Only use Drägersorb 800 Plus or
Drägersorb FREE!
Cleaning, disinfection and sterilization
C
soda lime.
WARNING
Risk of injury
Absorbent is caustic and is a strong eye, skin,
and respiratory tract irritant. Use care when
handling the absorbent to avoid spills.
NOTE
Do not use powdered lime, as a higher dust load
can impair the functionality of Primus Infinity
Empowered.
231