Changes In Peep; Changes In Tinsp; Changing Between Ventilation Modes - Dräger Primus Infinity Empowered Instructions For Use Manual

Anesthesia workstation
Hide thumbs Also See for Primus Infinity Empowered:
Table of Contents

Advertisement

The value for the ventilation parameter T
automatically turns yellow and is adjusted at the
same time. The ratio of inspiration to expiration I:E
remains constant.

Changes in PEEP

Depending on the configuration, the pressure limit
value P
can also be changed automatically
INSP
when changing the PEEP value, see page 176.
Pressure
C
B
For automatic P
adjustment:
INSP
1 Press the soft key PEEP (A), key lights up yel-
low.
2 Set and confirm the ventilation parameter PEEP
via the rotary knob.
The value for the ventilation parameter P
automatically turns yellow and is adjusted at the
same time.
If configured, the lower alarm limit for the airway
pressure P
(C) will be automatically changed
AW
when the PEEP value is changed.
Changes in T
INSP
T
may be reduced simultaneously if T
SLOPE
reduced.
Instructions for use Primus Infinity Empowered SW 4.5n

Changing between ventilation modes

(B)
INSP
When changing to a different ventilation mode, the
pre-settings are adopted or appropriately derived
from the parameters of the preceding mode.
Pressure
Parameters which are identical in both ventilation
modes are adopted directly (freq., T
P
, Trigger) (A).
PS
A
When changing from volume-controlled to pres-
sure-controlled ventilation:
The measured parameter PLAT is adopted as the
new value for P
When changing from pressure-controlled to vol-
ume-controlled ventilation:
The new tidal volume V
measured minute volume MV and set frequency
(B)
INSP
Freq.. Only the minute volume applied by the ven-
tilator is taken into account. Pressure supported
breathing strokes by the patient are disregarded.
When changing from automatic ventilation modes
to Pressure Support Mode (optional):
The set PEEP,
If
P
or Trigger were set to OFF, the last values
PS
used are adopted in Pressure Support mode. The
configured default settings are used in all other
is
INSP
cases.
B
A
INSP
(B).
INSP
is calculated based on the
T
P
, and Trigger are adopted.
PS
Operation
, PEEP,
117