Cleaning, disinfection and sterilization

Disassembly

Disconnecting from the mains
1 Shut down the device and all additional devices.
2 Remove the mains plug.

Observe before disassembly

Switch off the device and accessory devices
and remove their mains plugs.

Removing the sample line

Unscrew the sample line from the Y-piece and
the water trap on the front of the unit.
CAUTION
Risk of gas measurement failure and device fail-
ure
Disinfectants can damage the sample gas line
and the diaphragm of the water trap.
Sample gas lines are single-use articles and must
be replaced, not disinfected.
The sample line is a single-use article which must
be disposed of in accordance with the hospital's hy-
giene regulations.
216

Removing the water trap container

Pull the water trap out towards the front and
empty it, see page 82.
WARNING
Risk of gas measurement failure and device
failure
If the water trap is used longer than intended,
the diaphragm may become brittle and allow
water and bacteria to enter the measurement
system. Such contamination affects the gas
measurement which may fail as a result.
The water trap must be replaced after a serv-
ice life of four weeks.
When disposing of old water traps, follow the corre-
sponding instructions for use for the WaterLock wa-
ter trap or the Infinity ID WaterLock 2 and comply
with the hospital's hygiene requirements.