Disassembly - Dräger Julian Operating Instructions Manual

Anesthesia workstation
Hide thumbs Also See for Julian:
Table of Contents

Advertisement

Care

Disassembly

Removing the flow sensor
1 Unscrew expiratory port.
2 Remove flow sensor.
CAUTION !
Do not process flow sensor in cleaning and
disinfection equipment or steam-autoclave flow
sensor.
Do not use compressed air, a brush or similar
tools to clean flow sensor as this could possibly
damage the thin wires inside.
The flow sensor must be disinfected in 70 % ethanol
solution for approximately 1 hour.
Leave sensor to dry in air for at least 30 minutes,
otherwise the remaining alcohol can damage the
sensor when calibrating.
The flow sensor can be reused as long as it can be
calibrated successfully.
Removing the absorber and bellows
3 Turn absorber counterclockwise and pull downwards.
Empty the soda lime.
For disposal, refer to manufacturer's instructions.
4 Remove insert from the absorber.
WARNING !
Absorbent is caustic and is a strong irritant to
the eyes, skin, and respiratory tract. When
emptying absorber, take care not to spill its
contents.
5 Turn container of the bellows counterclockwise and
pull downwards.
6 Remove bellows.
Prepare absorber, bellows container and the bellows
for cleaning and disinfection in a parts washing/disin-
fection unit.
Removing the breathing system
Hold breathing system by its handle and pull it out.