Disassembly - Dräger Atlan A300 Instructions For Use Manual

Hide thumbs Also See for Atlan A300:
Table of Contents

Advertisement

Reprocessing
No.
3
4
5
6
7
8
9
10
11
12
13
14
15
3. If required, perform the surface disinfection for other components of the
anesthesia workstation, e.g., for a patient monitor.
4. After the product has been exposed to the disinfectant for the specified contact
time, remove residual disinfectant.
5. Wipe with a cloth dampened with water (at least drinking-water quality). Allow
the product to dry.
6. Check the product for visible soiling. Repeat steps 1 to 5 if necessary.
7. Check the product for visible damage and replace if necessary.
13.8

Disassembly

13.8.1
Preparation
1. Switch off the device and all devices connected to it.
2. Disconnect all power plugs.
13.8.2
Disassembling the water trap
The water trap is only present if the device is equipped with the integrated patient-
gas measurement module.
Prerequisite: The sample line has been removed.
Risk of infection
The water trap may contain infectious fluid.
► Proceed carefully when emptying and take protective measures if necessary.
► Follow the infection prevention policies and reprocessing regulations of the
230
Designation
Rotary knob
Control elements on the gas mixing unit
Work surface, side table (option)
Handles
Handles on drawers
Absorber
Circuit plug on bag elbow
Water trap on the patient-gas measurement module (PGM)
APL valve
Breathing system cover
O
+ key
2
Flow control valve on external O
Pressure reducers (on the rear of the device, not shown)
WARNING
health-care facility.
Instructions for use
flowmeter (option)
2
|
Atlan A300, A300 XL, A350, A350 XL SW 1.0n