Voluson® 730Expert Power Requirements
Table 2-3
Power Consumption nominal 1010 VA including all options.
Mains outlets: Mains socket ST1, ST2, ST3, ST4, ST5 for accessories.
All mains outlets are co-switched by the unit's mains switch via built-in isolation transformer.
Output voltage for: ST1 - ST5: 115V or 230V.
Modification of voltage setting only by an authorized service person!
CAUTION
The maximum power consumption of equipment (inclusive color video monitor) connected to
!! CAUTION:
these outlets must not exceed 350VA!
2-2-2-2

Inrush Current

Inrush current is not a factor to consider due to the inrush current limiting properties of the power
supplies.
2-2-2-3

Site Circuit Breaker

It is recommended that the branch circuit breaker for the machine be readily accessible.
POWER OUTAGE MAY OCCUR.
CAUTION
The Voluson® 730Expert requires a dedicated single branch circuit. To avoid circuit overload
!! CAUTION:
and possible loss of critical care equipment, make sure you DO NOT have any other equipment
operating on the same circuit.
2-2-2-4

Site Power Outlets

A dedicated AC power outlet must be within reach of the unit without extension cords. Other adequate
outlets for the external peripherals, medical and test equipment needed to support this unit must also
be present within 1 m (3.2 ft.) of the unit. Electrical installation must meet all current local, state, and
national electrical codes.
2-2-2-5

Main Power Plug

The Voluson® 730Expert is supplied with a main power plug, as standard.
In the event that the unit arrives without a power plug, or with the wrong plug, contact your GE dealer.
When necessary, the installation engineer will supply the appropriate power plug to meet the applicable
local regulations.
3
DRAFT (A
Electrical Specifications for Voluson® 730Expert
Voltage
Tolerances
100 VAC
110 VAC
115 VAC
130 VAC
230 VAC
240 VAC
Chapter 2 - Site Preparation
29, 2008)
PRIL
Current
±10%
10.10 A
±10%
9.20 A
±10%
8.80 A
±10%
7.80 A
±10%
4.40 A
±10%
4.20 A
V
® 730E
OLUSON
XPERT
S
ERVICE
Frequency
50, 60 Hz (±2%)
50, 60 Hz (±2%)
50, 60 Hz (±2%)
50, 60 Hz (±2%)
50, 60 Hz (±2%)
50, 60 Hz (±2%)
(BT03)
M
ANUAL
2-3