NI 5402/5406 Calibration Procedure
Figure 5 shows the programming flow for the optional calibration.
Self-Calibrate
Verify
Meets
Calibration
Test Limits?
Yes
Results
Within
Published Specs
(Adjustment
Optional)
Adjust
No
Anyway?
Figure 5. Optional Calibration Programming Flow
Document
Pre-Adjustment
Results
No
Meets
Published
Specifications?
(Cal Dates and
Yes
Temperatures
Updated)
Yes
Calibration
Test Limits?
98
No
Adjust
Document
Verify
Post-Adjustment
Results
Meets
No
Review
Verification/
Adjustment
Procedure or
Return Device