© National Instruments Corporation
Adjusting the Main Path Pre-Amplifier Gain
1.
Set the main DAC value by calling the niFgen Write Binary 16 Analog
Static Value VI.
LabVIEW Block Diagram
2.
Set the offset DAC value by calling the niFgen Property Node and
selecting Calibration»Offset DAC Value.
LabVIEW Block Diagram
67
C/C++
Function Call
Call
niFgen_Write
Binary16Analog
StaticValue
using the following
parameters:
vi: The session handle
returned from
niFgen_InitExtCal
channelName:
"0"
value:
0
C/C++
Function Call
Call
niFgen_Set
AttributeViInt32
using the following
parameters:
vi: The session handle
returned from
niFgen_InitExtCal
channelName:
"0"
attributeID:
NIFGEN_ATTR_OFFSET_
DAC_VALUE
value:
32,000
NI 5402/5406 Calibration Procedure
.
.