NI 5402/5406 Calibration Procedure
6.
Set the post-amplifier attenuation by calling the niFgen Property Node
and selecting Calibration»Post-Amplifier Attenuation.
LabVIEW Block Diagram
7.
Set the amplitude by calling the niFgen Property Node and selecting
Standard Function Output»Amplitude.
LabVIEW Block Diagram
18
C/C++
Function Call
Call
niFgen_Set
AttributeViReal64
using the following
parameters:
vi: The session handle
returned from
.
niFgen_init
channelName:
"0"
attributeID:
NIFGEN_
ATTR_POST_AMPLIFIER
_ATTENUATION
value: The
Post-Amplifier
Attenuation value listed in
Table 3 for the current
iteration
C/C++
Function Call
Call
niFgen_Set
AttributeViReal64
using the following
parameters:
vi: The session handle
returned from
.
niFgen_init
channelName:
"0"
attributeID:
NIFGEN_
ATTR_FUNC_AMPLITUDE
value: The Amplitude
value listed in Table 3 for
the current iteration