National Instruments NI 5402 Calibration Procedure page 82

Hide thumbs Also See for NI 5402:
Table of Contents

Advertisement

NI 5402/5406 Calibration Procedure
4.
Set the flatness correction state by calling the niFgen Property Node
and selecting Calibration»Flatness Correction Enabled.
LabVIEW Block Diagram
5.
Configure the waveform by calling the niFgen Configure Standard
Waveform VI.
LabVIEW Block Diagram
82
C/C++
Function Call
Call
niFgen_Set
AttributeViBoolean
using the following
parameters:
vi: The session handle
returned from
niFgen_
InitExtCal
channelName:
"0"
attributeID:
NIFGEN_
ATTR_FLATNESS_
CORRECTION_ENABLED
value:
VI_FALSE
C/C++
Function Call
Call
niFgen_
ConfigureStandard
using the
Waveform
following parameters:
vi: The session handle
returned from
niFgen_InitExtCal
channelName:
"0"
waveform:
NIFGEN_VAL_WFM_SINE
frequency:
50,000
amplitude:
2.00
startPhase:
0
dcOffset:
0
.
ni.com